Model Number 500101-001 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Death (1802); Ischemia (1942); Liver Damage/Dysfunction (1954); Respiratory Failure (2484); Multiple Organ Failure (3261)
|
Event Date 01/18/2018 |
Event Type
Death
|
Manufacturer Narrative
|
The syncardia 70 cc temporary total artificial heart (tah-t) is an implantable pulsatile biventricular replacement device that replaces a patient's native ventricles and valves and pumps blood to both the pulmonary and systemic circulation systems.The syncardia 70 cc tah-t is indicated for use as a bridge to transplantation in cardiac transplant-eligible candidates at risk of imminent death from biventricular failure.The syncardia tah-t system is intended for use inside and outside the hospital.The tah-t will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4) initial.
|
|
Event Description
|
The customer, a syncardia certified hospital, reported that the patient expired on (b)(6) 2018 from acute hepatic failure and respiratory failure.Additional customer comments and observations reported on the outcome form: evidence of ischemic injury to multiple organs including the presence of geographic appearing infarcts within bilateral testicular parenchyma and spleen; the artificial heart was in place without significant associated hemorrhage or hematoma.The customer also reported that the tah-t was explanted and an autopsy was performed on (b)(6) 2018.No further information has been provided at this time.
|
|
Manufacturer Narrative
|
Visual inspection, valve evaluation, thrombus formation evaluation, and a stroke volume evaluation were performed during the tah-t explant analysis.No anomalies or abnormalities were observed.This issue will continue to be monitored and trended as part of the customer experience process.Syncarida has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|
|
Search Alerts/Recalls
|