STRYKER ENDOSCOPY-SAN JOSE PKG, 5MM PEEK HANDLE, 45CM; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
|
Back to Search Results |
|
Catalog Number 0250181164 |
Device Problem
Device Packaging Compromised (2916)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/06/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Alleged failure: bent shaft|burn on insulation at shaft.Confirmed failure: bent shaft, burn on insulation at shaft.Probable root cause: poor autoclave reliability, incorrect sterilization/reprocessing procedure, handling procedures, contact forces, product used beyond defined useful life.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.The device manufacture date is not known.(b)(4).
|
|
Event Description
|
It was reported that the insulation had been compromised.
|
|
Manufacturer Narrative
|
The product was returned for investigation and the reported failure mode was confirmed.Alleged failure: bent shaft|burn on insulation at shaft.Confirmed failure: bent shaft, burn on insulation at shaft.Probable root cause: poor autoclave reliability, incorrect sterilization/reprocessing procedure, handling procedures, contact forces, product used beyond defined useful life.The failure mode will be monitored for future reoccurrence.The device manufacture date is not known.Gtin: (b)(4).
|
|
Event Description
|
It was reported that the insulation had been compromised.
|
|