MAUDE Adverse Event Report: ACUMED LLC ARH SLIDE-LOC¿ STANDARD STEM 11MM; PROSTHESIS, ELBOW, HEMI-, RADIAL, STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACUMED LLC ARH SLIDE-LOC¿ STANDARD STEM 11MM; PROSTHESIS, ELBOW, HEMI-, RADIAL, STEM Back to Search Results
Model Number 5001-0111N-S
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
The head, neck, and stem was already assembled together when received.The components were separated to obtain the part and batch numbers.The laser marks between the head and neck were aligned.There were some marks/damage on the neck flange and stem platform.The unlocking torque was calculated to be less than the required (as manufactured) specification.The diameter of the stem measured to be within manufacturing specifications.Additional mdrs associated with this event: 3025141-2018-00115: head.3025141-2018-00116: neck.
 
Event Description
An arh slide-loc radial head replacement system was explanted on (b)(6) 2018 for unknown reasons.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARH SLIDE-LOC¿ STANDARD STEM 11MM
Type of Device
PROSTHESIS, ELBOW, HEMI-, RADIAL, STEM
Manufacturer (Section D)
ACUMED LLC
5885 ne cornelius pass road
hillsboro OR 97124 9432
Manufacturer (Section G)
ACUMED LLC
5885 ne cornelius pass road
hillsboro OR 97124 9432
Manufacturer Contact
micki lehman
5885 ne cornelius pass road
hillsboro, OR 97124
8886279957
MDR Report Key7314410
MDR Text Key101562074
Report Number3025141-2018-00117
Device Sequence Number1
Product Code KWI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131845
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 02/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/05/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date07/31/2022
Device Model Number5001-0111N-S
Device Catalogue Number5001-0111N-S
Device Lot Number359467
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/22/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/12/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-