The replaced inspiratory module was returned for failure investigation.Upon receipt a visual inspection revealed that the module was returned without the water trap and the 02 sensor was broken.The unit was severely damaged with liquid contamination.There were splash marks evident on the module.When disassembling the unit to remove the air flow sensor for inspection liquid spilled out onto the working surface.Once the flow sensor was removed it was noted that the port into which the flow sensor tubing is seated was full of water.Because of the extensive damage and contamination, it would not be safe to attach the returned inspiratory module and power on the ventilator.The preventative maintenance section of the 840 operators and technical reference manual gives instructions on preventative maintenance procedures and frequency for expiratory and inspiratory limbs, bacteria filters, water traps and drain bags.If information is provided in the future, a supplemental report will be issued.
|