• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS MEDICAL SOLUTIONS USA, INC. ACUSON SC2000 ULTRASOUND SYSTEM; ULTRASOUND DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS MEDICAL SOLUTIONS USA, INC. ACUSON SC2000 ULTRASOUND SYSTEM; ULTRASOUND DEVICE Back to Search Results
Model Number ACUSON SC2000
Device Problems Computer Software Problem (1112); Device Operational Issue (2914)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/22/2018
Event Type  malfunction  
Manufacturer Narrative
The device referenced in this report has not been returned to siemens for evaluation.If additional information is received or if the device is returned at a later date, this report will be supplemented.
 
Event Description
It was reported that the ultrasound system shutdown intermittently in the middle of an unspecified procedure.The user rebooted the ultrasound system but the problem persisted.Another system was brought in to complete the procedure.There was no loss of data and there was no patient adverse event reported.No additional information was provided.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the new date received by manufacturer (see section g4); update the device available for evaluation (see d10), provide the type of report (see section g7); provide the type of reportable event (see section h1); provide the type of follow-up (see section h2); update the event problem and evaluation codes (see section h6); and provide additional manufacturer narrative (see section h10).In this case, the customer service engineer (cse) replaced the power supply module (psm) and reloaded the software.The psm associated with the event was returned to the supplier, who performed tests on the unit.The supplier stated that the psm unit passed all tests, and the reported issue could not be reproduced by them.Review of the log files provided by the cse found multiple instances of improper shut down sequences by the customer.Based on the investigation, engineer stated that it is likely that the improper shut down sequences done by the customer corrupted the hard disk, which caused the reported symptoms.The acuson sc2000 system instructions for use state: warning: to prevent the loss of data that results from unplugging the system from the power outlet before the system has completed its shutdown sequence, follow the system shutdown instructions in the system setup chapter in the instructions for use manual.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUSON SC2000 ULTRASOUND SYSTEM
Type of Device
ULTRASOUND DEVICE
Manufacturer (Section D)
SIEMENS MEDICAL SOLUTIONS USA, INC.
685 east middlefield road
mountain view CA 94043 4050
MDR Report Key7317717
MDR Text Key101854668
Report Number3009498591-2018-00010
Device Sequence Number1
Product Code IYN
Combination Product (y/n)N
PMA/PMN Number
K113179
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 10/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberACUSON SC2000
Device Catalogue Number10433816
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2018
Date Manufacturer Received10/15/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-