• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION PENTA 3MM LEAD, 60 CM; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION PENTA 3MM LEAD, 60 CM; SCS LEAD Back to Search Results
Model Number 3228
Device Problem Detachment Of Device Component (1104)
Patient Problem Inadequate Pain Relief (2388)
Event Date 01/26/2018
Event Type  Injury  
Manufacturer Narrative
The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient experienced inadequate pain relief due to intermittent decrease of stimulation for the past 6 months.Reprogramming was unable to achieve complete pain pattern coverage as a result, a physician consult was scheduled as the next course of action.Follow-up identified surgical intervention was undertaken on (b)(6) 2018 to address the issue.During the procedure, an electrode was found in the subcutaneous tissue.In turn, the electrode was removed from the tissue as well as the lead was replaced with a new model.Postoperatively, effective stimulation was restored and the issue resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTA 3MM LEAD, 60 CM
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7319255
MDR Text Key101689345
Report Number1627487-2018-02253
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2017
Device Model Number3228
Device Lot Number5069211
Other Device ID Number05414734406246
Was Device Available for Evaluation? No
Date Manufacturer Received02/08/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/30/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age40 YR
-
-