Model Number 10-60-00 |
Device Problems
Decreased Pump Speed (1500); Pumping Stopped (1503)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/06/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient identifier not provided.This information will be provided in a supplemental report if made available.Livanova (b)(4) manufactures the s3 roller pump.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
|
Event Description
|
Livanova (b)(4) received a report that a s3 roller pump became slow and finally stopped during a procedure without an alarm.The customer has to use the hand-crank for two minutes.After that the pump switched on and worked without further issues.There was no report of patient injury.
|
|
Manufacturer Narrative
|
Through follow up communication, livanova (b)(4) learned, that the affected equipment has been taken out of service.
|
|