• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S3 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S3 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-60-00
Device Problems Decreased Pump Speed (1500); Pumping Stopped (1503)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/06/2018
Event Type  malfunction  
Manufacturer Narrative
Patient identifier not provided.This information will be provided in a supplemental report if made available.Livanova (b)(4) manufactures the s3 roller pump.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
 
Event Description
Livanova (b)(4) received a report that a s3 roller pump became slow and finally stopped during a procedure without an alarm.The customer has to use the hand-crank for two minutes.After that the pump switched on and worked without further issues.There was no report of patient injury.
 
Manufacturer Narrative
Through follow up communication, livanova (b)(4) learned, that the affected equipment has been taken out of service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S3 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
MDR Report Key7319947
MDR Text Key101731817
Report Number9611109-2018-00182
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K950990
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-60-00
Was the Report Sent to FDA? No
Date Manufacturer Received09/22/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age80 YR
Patient Weight53
-
-