• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SAFETY-LOK¿ PRE-ATTACHED BLOOD COLLECTION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SAFETY-LOK¿ PRE-ATTACHED BLOOD COLLECTION SET Back to Search Results
Catalog Number 368653
Device Problems Fluid/Blood Leak (1250); Hole In Material (1293)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/07/2018
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone: (b)(6).Initial reporter email: (b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd vacutainer® safety-lok¿ pre-attached blood collection set leaked blood into the hands of the phlebotomist from a hole in the tubing.No medical intervention or serious injury reported.Found during use.
 
Manufacturer Narrative
Investigation summary: bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for a small cut in the tubing with the incident lot was observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd has initiated further investigation relating to this issue through a capa.The investigation is still on-going and improvements will be made as the potential causes of this issue are identified.Investigation conclusion: based on evaluation of the customer samples, the customer¿s indicated failure mode for a small cut in the tubing with the incident lot was observed.Further investigation has been initiated through a capa.The investigation is still on-going and improvements will be made as the potential causes are identified.Root cause description: a capa has been initiated to document further investigation and root cause analysis relating to this issue.The investigation is currently on-going and will be updated as the potential root cause(s) are identified.Rationale: based on an assessment of severity and frequency, it was determined that a capa is required at this time in order to determine the root cause associated with this issue and implement corrective actions.The investigation is currently on-going and further improvements will be made as the potential root cause(s) are identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SAFETY-LOK¿ PRE-ATTACHED BLOOD COLLECTION SET
Type of Device
BLOOD COLLECTION SET
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
MDR Report Key7322738
MDR Text Key101979199
Report Number1024879-2018-00075
Device Sequence Number1
Product Code JKA
UDI-Device Identifier50382903686532
UDI-Public50382903686532
Combination Product (y/n)N
PMA/PMN Number
K991088
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Phlebotomist
Device Expiration Date09/30/2019
Device Catalogue Number368653
Device Lot Number7275864
Date Manufacturer Received02/16/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-