MAUDE Adverse Event Report: BARD ACCESS SYSTEMS PER-Q-CATH PLUS 2F SINGLE-LUMEN PICC WITH STIFFENING STYLET BASIC TRAY (30 CM); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS PER-Q-CATH PLUS 2F SINGLE-LUMEN PICC WITH STIFFENING STYLET BASIC TRAY (30 CM); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number 4132105
Device Problems Migration or Expulsion of Device (1395); Device-Device Incompatibility (2919)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/09/2018
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer for evaluation, as the device was discarded after the event occurred.A lot history review (lhr) of rebt0803 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (rebt0803) have been reported from the same (b)(6) facility.
 
Event Description
During catheter placement, in all the veins that were punctured, the catheter was migrating when reaching around 9 centimeters.By the time the catheter was withdrawal, it was noticed that the guide wire had perforated the catheter.However the correct technique was done by keeping one centimeter drawn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PER-Q-CATH PLUS 2F SINGLE-LUMEN PICC WITH STIFFENING STYLET BASIC TRAY (30 CM)
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas
MX  
Manufacturer Contact
shelly gilbert
605 n. 5600 w.
salt lake city, UT 84116
8015225640
MDR Report Key7324386
MDR Text Key102000155
Report Number3006260740-2018-00399
Device Sequence Number1
Product Code LJS
UDI-Device Identifier00801741074837
UDI-Public(01)00801741074837
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K954104
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 03/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2022
Device Model Number4132105
Device Catalogue Number4132105
Device Lot NumberREBT0803
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Hospital
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-