• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 195240
Device Problem Failure To Adhere Or Bond (1031)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/13/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4), pad, 3194 - adhesive.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the level sensor pads would not stick.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Updated block: concomitant medical products.Per the manufacturer's territory manager, the pads were not mounted on a seam and the user facility waited to the recommended five minutes before attaching the level sensor.
 
Manufacturer Narrative
(b)(4).The reported complaint could not be confirmed.Additional parts returned were determined to be applicable to this complaint.During laboratory analysis, the product surveillance technician (pst) was not able to duplicate the reported issue.The mounting pads stay adhered to oxygenator throughout the three hour testing period.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.This complaint is related to (b)(4) / medwatch #1828100-2018-00068.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7326768
MDR Text Key102158020
Report Number1828100-2018-00115
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/20/2019
Device Model Number195240
Device Catalogue Number195240
Device Lot Number847342
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2018
Date Manufacturer Received05/04/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MEDTRONIC FUSION RESERVOIR
-
-