The 3500cp-g mixer, serial number (b)(4), was returned due to the reported issue of the mixer unit failing "to oxygenate and flow co2"during patient use.No patient injury.The device did not oxygenate and flow gas due to improper usage by the user.The 3500cp-g mixer is a "precision gas mixing device intended for use with medical breathing air and medical breathing oxygen only" per the user's manual.Also stated in the user's manual is the "user of the sechirst air/oxygen mixer shall have sole responsibility for any malfunction, which results from improper usage.The failures observed during full testing per doc.(b)(4) recorded in the product evaluation section of this report are due to the failure of the device not being overhauled per the manufacturer's recommended interval of every 2 years.A dhr (device history review) was performed.The 3500cp-g mixer, serial number (b)(4), was manufactured on 6/11/2015.There is no indication that there were any relevant discrepancies during manufacturing.Device history record (dhr) did not find any non-conformance that could cause or contribute to the reported issue.
|