The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The needle was extended from the outer tube.The outer tube was kinked.The manufacturing record was reviewed and found no irregularities.This type of the event is most likely related to the operator's technique.Based on the past similar cases, it was known that the frictional resistance between the outer tube and the needle tube increased due to the kinks, and then the needle could not be retracted into the outer tube.Also, the kinks of the outer tube occurred due to applying bending load to the outer tube when the subject device was inserted into the endoscope, it was taken out from the sterile bag, or pre-use inspection was performed.The instruction manual of the device has already warned as follows; insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.Stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.
|