BECTON, DICKINSON & CO., (BD) BD VACUTAINER® ADAPTER WITH LUER ADAPTER; BLOOD COLLECTION SET Back to Search Results
Catalog Number 367300
Device Problem Hole In Material (1293)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/26/2018
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd vacutainer® adapter with leur adapter had a hole in the tip of the shield.Found before use.No reports of serious injury or medical intervention noted.
 
Manufacturer Narrative
Bd received samples and photos from the customer facility for investigation.The photos were evaluated and a hole in the top of the iv shield with the incident lot was observed.Additionally, evaluation of the customer samples was performed and a hole in the top of the iv shield was observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: based on evaluation of the customer photos and samples, the customer¿s indicated failure mode for a hole in the top of the iv shield with the incident lot was observed.Based on the investigation, the most likely root cause was determined to be related to a manufacturing issue.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® ADAPTER WITH LUER ADAPTER
Type of Device
BLOOD COLLECTION SET
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7331927
MDR Text Key102317132
Report Number1024879-2018-00081
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K921520
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/31/2020
Device Catalogue Number367300
Device Lot Number7139695
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received02/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-