Catalog Number QD11 |
Device Problems
Overheating of Device (1437); Device Inoperable (1663)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/01/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure it was observed that the angle attachment device was heating up and was inoperable.It was not reported if there was a delay in the procedure due to the event.It was not reported if a spare device was available for use.There was patient involvement reported.There were no patient or user injuries reported.It was reported there was no medical intervention or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2018.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the locking components and bearings were damaged.The device also failed pretest for thimble lock operation, nose tube assembly and cutter insertion.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
Upon further investigation of the device evaluation, it was determined that the device controls were blocked, the bearing and sleeve were damaged and the glue joint to sleeve was loose.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|