Catalog Number 121932058
Device Problems Disassembly (1168); Device Dislodged or Dislocated (2923)
Patient Problems Not Applicable (3189); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Complaint description: depuy17-30.  it was occured a dislocation of the pinnacle marathon pe neutral liner 32idx58od that was implanted in 2011.The event was not linked to any trauma.It was necessary a new intervention.
 
Manufacturer Narrative
Product complaint: (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: updated february 27, 2019.The device was returned for examination.Examination of the returned acetabular liner finds sufficient evidence to confirm the reported disassociation event.The investigation has not identified product error.The manufacturing records have been reviewed and no related deviations or anomalies were identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR NEUT 32IDX58OD
Type of Device
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7333620
MDR Text Key102199987
Report Number1818910-2018-54931
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295013938
UDI-Public10603295013938
Combination Product (y/n)N
PMA/PMN Number
K001534
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup,Followup
Report Date 08/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number121932058
Device Lot NumberE15RY1000
Was Device Available for Evaluation? No
Date Returned to Manufacturer10/09/2017
Date Manufacturer Received05/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-