ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE CONN./EXT. DL; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
|
Back to Search Results |
|
Model Number LIBERTY CYCLER SET, |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
Inflammation (1932)
|
Event Date 02/21/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of the plant¿s investigation.
|
|
Event Description
|
Peritoneal dialysis patient reported observing a fluid leak on the floor and in the liberty select cassette compartment.The cycler did alarm but the patient could not confirm the type of alarm that was encountered.Follow up with the patients peritoneal dialysis nurse indicates that the patient did have abdominal inflammation but was not diagnosed with any infection.The patient was prescribed prophylactic antibiotics, ip vancomycin 3 times a week and oral keflex.The liberty cassette was discarded.
|
|
Manufacturer Narrative
|
Correction: weight.
|
|
Event Description
|
".".
|
|
Manufacturer Narrative
|
The alleged event is not confirmed.The device was not returned to the manufacturing plant for investigation.All companion samples have been sold and distributed.An investigation of the product history records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, the product history review confirmed the labeling, material, and process controls were within specification.
|
|
Event Description
|
".".
|
|