• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEURSCHLAND S3 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEURSCHLAND S3 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-40-00
Device Problems Device Displays Incorrect Message (2591); No Flow (2991)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/26/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.The gas blender system 25-40-00 is not distributed in the usa, but it is similar to gas blender system 25-40-45, which is distributed in the usa (510(k) number: k052601).This event was initially considered to be non-reportable.However, after additional evaluation, livanova (b)(4) has decided to reclassify this event as reportable.This report is being filed as part of a retrospective review conducted in response to this new decision.Livanova (b)(4) manufactures the s3 gas blender system.The incident occurred in (b)(6).The device was sent back to the manufacturer for investigation and the complaint was confirmed.The root cause could be traced back to a defect in the front panel and was therefore replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Event Description
Livanova (b)(4) received a report that a s3 gas blender system displayed an error message during procedure.After the error message appeared the gas flow stopped.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S3 GAS BLENDER SYSTEM
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEURSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan caesar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7333842
MDR Text Key102223659
Report Number9611109-2018-00205
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeNL
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 03/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25-40-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/29/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-