The customer stated that they received questionably high results for an unspecified number of patient samples tested for nh3l ammonia (nh3) on a cobas 6000 c (501) module - c501.Samples will repeat in normal ranges.The customer provided data for one patient sample which had an erroneous nh3 result that was reported outside of the laboratory.The sample initially resulted as 193.5 ug/l.The value was reported outside of the laboratory to the doctor, who asked for a rerun of the sample since the value was not compatible with the patient's clinical history.The sample was repeated on the same analyzer, resulting as 33.1 ug/dl.The sample was also repeated on a second c501 analyzer, resulting as 30.6 ug/l.No adverse events were alleged to have occurred with the patient.The nh3 reagent lot number was 273314, with an expiration date of 31-mar-2019.The last calibration performed on (b)(6) 2018 was flagged with an error.Quality control recovery was not stable, but was within range on the date of the event.Several abnormal probe aspiration alarms occurred on (b)(6) 2018.
|