BAXTER HEALTHCARE - MOUNTAIN HOME MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4482 |
Device Problem
Fracture (1260)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/18/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).The device was received for evaluation.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Leak testing, clear passage testing and clamp function testing were performed with no issues noted.A visual inspection was performed with the naked eye and revealed a cracked main body of the minicap transfer set.The cause of the crack could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
During evaluation of a minicap transfer set, it was found that the main body was cracked.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|