ALCON RESEARCH, LTD. - ALCON PRECISION DEVICE V-LANCE KNIVES; KNIFE, OPHTHALMIC Back to Search Results
Catalog Number 8065912001
Device Problems Bent (1059); Dull, Blunt (2407)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/22/2018
Event Type  malfunction  
Manufacturer Narrative
A sample is available that has not yet been received at manufacturing for evaluation.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Additional information is confirmed to be unavailable for this report.The manufacturer internal reference number is: (b)(4).
 
Event Description
A doctor reported that a knife blade was dull and exhibited a bent tip during surgery.There was no impact to the patient.Additional information is confirmed to be unavailable.
 
Manufacturer Narrative
One opened knife sample with foam protector was received in a blister package, not related to the complaint, for the report of a dull blade and bent tip.The returned sample was visually inspected and was found to be nonconforming with a damaged tip and damaged cutting edges.Penetration and sharpness testing could not be performed due to the damaged condition of the sample.No lot number was identified with this complaint therefore, lot history and complaint history reviews could not be conducted.The exact root cause could not be determined from the investigation performed.The damage to the returned sample is consistent with damage that can occur when the blade contacts a hard surface such as the protective blade tray when the product is improperly removed or inserted after use, from improper handling or from contact with another instrument during surgery or set-up.The exact root cause for the damaged knife sample is unknown therefore, specific action with regards to this complaint cannot be taken.All knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.Any nonconformance, such as the damaged tip and damaged cutting edge exhibited on the returned opened sample, is removed from the lot and scrapped.Functional penetration testing is performed and monitored during the finishing process to ensure the sharpness of the product.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-LANCE KNIVES
Type of Device
KNIFE, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LTD. - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LTD. - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7337727
MDR Text Key102463687
Report Number2523835-2018-00129
Device Sequence Number1
Product Code HNN
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065912001
Device Lot NumberASKU
Other Device ID Number380659120015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/24/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
-
-