Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was kinked approximately 7.0 cm from the proximal end.The pusher assembly was fractured approximately 5.0 cm from the proximal end.The embolization coil was intact with its pusher assembly.Conclusions: evaluation of the returned device confirmed that the ruby coil¿s pusher assembly was kinked.If the device is forcefully advanced against resistance, damage such as this may occur.Further evaluation of the returned device revealed that the ruby coil pusher assembly was fractured.This damage was likely incidental and may have occurred while packaging the device for return to penumbra.The lantern referred to in the complaint was not returned for evaluation.Penumbra coils and catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a coil embolization procedure using ruby coils.During the procedure, the pusher assembly of a ruby coil became kinked as the physician attempted to advance the ruby coil into a lantern delivery microcatheter (lantern).The ruby coil was therefore removed, and the procedure was completed using a new ruby coil.There was no report of an adverse effect to the patient.
|