Model Number WCD 4000
Device Problems Leak/Splash (1354); Patient-Device Incompatibility (2682)
Patient Problem Skin Irritation (2076)
Event Date 02/20/2018
Event Type  Injury  
Manufacturer Narrative
Biocompatibility testing to iso 10993 was successfully completed on skin-contacting surfaces of the lifevest device as well as the blue(tm) defibrillation gel.
 
Event Description
A distributor contacted zoll to report that a patient had a red, tender area under the therapy electrode pads.The patient reported that gel was leaking, causing the irritation.The patient's physician advised him to use a cream.Follow-up indicated that the irritation had improved.Upon evaluation of the electrode belt at the distributor, there was gel leaking from all therapy electrodes.The belt's ecg acquisition and pulse delivery circuitry were verified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
gabrielle salazar
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key7338680
MDR Text Key102382508
Report Number3008642652-2018-02550
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005005
UDI-Public00855778005005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 03/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/07/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age39 YR
-
-