ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2272
Device Problems No Device Output (1435); Pacemaker Found in Back-Up Mode (1440); Telemetry Discrepancy (1629)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/15/2018
Event Type  Injury  
Event Description
It was reported that the patient presented in the operating room for an implant procedure.When connecting the pulse generator to the lead, pacing could not be observed immediately.While performing lead impedance test, rf transmission with the device could not be established.During interrogation by wand, the device was noted to be in back-up.The device was explanted and replaced.There were no adverse consequences to the patient.
 
Manufacturer Narrative
The reported field event of backup vvi (bvvi) was confirmed in the laboratory and it was due to an unknown por.Device was tested on the bench and no anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7339641
MDR Text Key102411542
Report Number2017865-2018-03702
Device Sequence Number1
Product Code DXY
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2019
Device Model NumberPM2272
Device Lot NumberP000054746
Other Device ID Number05414734509589
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/22/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
-
-