Catalog Number EMAX2PLUS |
Device Problems
Overheating of Device (1437); Device Operates Differently Than Expected (2913); Noise, Audible (3273)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/01/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the temperature of the device was above specification.Therefore, reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the motor device was making an abnormal noise and had smoke coming out of it.It was not reported if there was a delay in the procedure due to the event, or if a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported, however, it was noted that the event occurred in 2018.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|