Catalog Number EMAX2PLUS
Device Problems Overheating of Device (1437); Device Operates Differently Than Expected (2913); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the temperature of the device was above specification.Therefore, reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the motor device was making an abnormal noise and had smoke coming out of it.It was not reported if there was a delay in the procedure due to the event, or if a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported, however, it was noted that the event occurred in 2018.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMAX 2 PLUS MOTOR
Type of Device
MOTOR, DRILL, ELECTRIC - HANDPIECE
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key7340745
MDR Text Key102457916
Report Number1045834-2018-50522
Device Sequence Number1
Product Code HBC
UDI-Device Identifier845384002055
UDI-Public(01)845384002055(11)160909
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 02/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEMAX2PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2018
Date Manufacturer Received02/16/2018
Date Device Manufactured09/09/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-