• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER; NRY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER; NRY Back to Search Results
Catalog Number 5MAXACE068
Device Problems Break (1069); Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311)
Patient Problem No Patient Involvement (2645)
Event Date 02/12/2018
Event Type  malfunction  
Manufacturer Narrative
Results: the penumbra system ace 68 reperfusion catheter (ace68) was fractured just distal to the strain relief, approximately 3.5 cm from the hub.Conclusions: evaluation of the returned device confirmed it was fractured just distal to the strain relief.This damage may have occurred due to forceful handling of the ace68 at extreme angles during removal from the packaging hoop.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
During preparation for a medical procedure, the scrub technologist found that the penumbra system ace 68 reperfusion catheter (ace68) was broken at the distal to the proximal hub upon removal from the packaging.The damage to the ace68 was found prior to use.Therefore the ace68 was not used in the procedure.The procedure was completed using a new ace68.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER
Type of Device
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key7342922
MDR Text Key102595710
Report Number3005168196-2018-00579
Device Sequence Number1
Product Code NRY
UDI-Device Identifier00814548016535
UDI-Public00814548016535
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K161640
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/01/2005,02/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2020
Device Catalogue Number5MAXACE068
Device Lot NumberF80301
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received01/14/2005
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-