Model Number 518-062
Device Problem Detachment Of Device Component (1104)
Patient Problems Death (1802); Exsanguination (1841); Low Blood Pressure/ Hypotension (1914)
Event Date 02/16/2018
Event Type  Death  
Manufacturer Narrative
Device lot number and expiration date are not available from the facility.The device was discarded by the user.Device manufacture date is dependent on the device lot number, thus is unavailable.
 
Event Description
It was reported that during a lead extraction procedure to remove 2 cardiac leads (rv, ra), an injury occurred.Both leads were prepped with lld devices.Extraction began with the rv lead using a glidelight laser sheath, and binding was experienced on the superior vena cava (svc) coil in the innominate vein.Efforts were then switched to the ra lead to create space.Smooth progress was made halfway down the atrial lead until it came loose with the use of traction force.Blood pressure then dropped and the patient was prepped for surgical intervention.A sternotomy was promptly performed and a 5cm posterior svc tear was identified.The injury could not be repaired and the patient did not survive.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
LLD
Manufacturer (Section D)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
Manufacturer Contact
nick mcnabb
9965 federal drive
colorado springs, CO 80921
719447-257
MDR Report Key7343075
MDR Text Key102535906
Report Number1721279-2018-00033
Device Sequence Number1
Product Code DRB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142116
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model Number518-062
Device Catalogue Number518-062
Device Lot NumberUNAVAILABLE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM; SPECTRANETICS GLIDELIGHT LASER SHEATH
Patient Outcome(s) Death;
Patient Age64 YR
Patient Weight122
-
-