FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR INTERFACE; CAT Back to Search Results
Model Number OPT318
Device Problem Device Issue (2379)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/11/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The optiflow junior cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and paediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.Method: the complaint opt318 cannula was not received for evaluation.We have based our investigation on a photograph provided by the customer.Results: photographic inspection revealed that one of the tubes was stretched and damaged at the swivel grip tube joint.Although the tubing was stretched, the metal spring was still intact and attached to the swivel grip.Conclusion: the damage to the flexitube was most likely caused by an excessive pulling force.All optiflow junior cannulae are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are currently taken hourly from each run and pull tested to check glue joint strength at the cannula/tube joint, as well as the swivel grip joint.If there are any failures the entire batch is put aside for further investigation.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior nasal cannula.They also state the following: do not stretch or crush tube.Ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.Appropriate monitoring must be used at all times.
 
Event Description
A hospital in (b)(6) reported that an opt318 optiflow junior nasal cannula was found broken on removal from the patient.The cannula was exchanged for another one and there was no further patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW JUNIOR INTERFACE
Type of Device
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
sharon kuruvilla
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key7344188
MDR Text Key102852353
Report Number9611451-2018-00201
Device Sequence Number1
Product Code CAT
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other
Type of Report Initial
Report Date 01/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT318
Device Catalogue NumberOPT318
Device Lot Number2100275160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/27/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-