MAUDE Adverse Event Report: AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER Back to Search Results
Catalog Number SGC0301
Device Problems Break (1069); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/21/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event Description
This is filed to report a cable break.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.The steerable guide catheter (sgc) was advanced in the anatomy while slightly turning the minus knob.A noise was heard and the feel of the device changed confirming that there was a cable break.The sgc was removed from the patient.A new sgc was prepped and used for the procedure.One clip was implanted successfully and the final mr was a grade of 1.No adverse patient effects or clinically significant delay in the procedure were reported.No additional information was provided.
 
Manufacturer Narrative
Internal file number: (b)(4).Evaluation summary: all available information was investigated and the reported cable break was confirmed.The reported noise (device operates differently than expected) could not be replicated in a testing environment as it was likely a symptom of the cable break.A review of the lot history record revealed no manufacturing nonconformities.Additionally, a review of the complaint history identified no similar incidents reported from this lot.Based on the information reviewed, the noise was due to the cable break; however, a definitive cause for the cable break could not be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
STEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
MENLO PARK, CA REG# 3005070406
abbott vascular
26531 ynez rd.
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key7347100
MDR Text Key103053852
Report Number2024168-2018-01927
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/07/2018
Device Catalogue NumberSGC0301
Device Lot Number71206U209
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/09/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
Patient Weight82
-
-