Model Number G8
Device Problems Complete Blockage (1094); Device Displays Incorrect Message (2591)
Patient Problem No Information (3190)
Event Date 06/08/2016
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) went to the customer site to evaluate the problem and perform preventive maintenance (pm) on the g8 analyzer.Following, the fse performed calibration, ran quality controls and did precision testing on the g8 analyzer.Operation of the g8 analyzer was verified and placed back into use.Per the service report no large syringe error code occurred.It is believed that the sampling needle was clogged which prevented the syringe from pushing fluid out of the needle.Replacing of the clogged needle with a new one, the system was able to flush through the needle as expected.The most probable cause of the reported event is related to clogging of the specimen nozzle.Review of the g8 variant analysis operator's manual indicates the error message "706 syringe-l error is an operation error in syringe-l.Counter measure: inspect syringe-l.Execute smp.Reset.Error level 2: enter wash state.Alarm level 1: beep for 30 sec and turn on error led & patolight® (optional).Review of the tosoh automated glycohemoglobin analyzer g8 service manual for 706 syringe-l error addresses the following information: content: operation error in syringe-l monitoring method: unexpected home sensor and limiter sensor signals are monitored during syringe l operation cause: a) incorrect syringe l operation b) syl motor fault c) syl-u sensor fault d) syl-d sensor fault e) drv2 board fault f) asm board fault countermeasure: a) inspect syringe l b) inspect the syl motor c) inspect sensors d) inspect boards.(b)(4).This report is being submitted due to a retrospective review conducted under capa-(b)(4).
 
Event Description
On (b)(6) 2016 the customer called tosoh technical support to report getting a large syringe error code 706 during use of the g8 analyzer.The customer further reported that they already replaced the sampling needle for this issue.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting of patient results for hba1c.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court, ste 101
south san francisco, CA 94080
6506368123
MDR Report Key7347884
MDR Text Key103165518
Report Number8031673-2018-02226
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/18/2018
Distributor Facility Aware Date06/08/2016
Device Age1 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/18/2018
Date Manufacturer Received06/08/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-