Results: the cat6 was stretched and fractured approximately 124.0 cm from the hub.The distal segment of the cat6 was ovalized and kinked throughout its length.The proximal segment of the cat6 was kinked in multiple locations throughout its length.Conclusions: evaluation of the returned device revealed the cat6 was stretched and fractured in the distal shaft.It is possible for the cat6 to become pinned against patient anatomy or other devices used in the procedure.If the cat6 becomes pinned or otherwise restrained, and the guidewire or cat6 is forcefully retracted against resistance, stretching and fracture damage may occur.Further evaluation revealed the distal segment of the cat6 was kinked and ovalized throughout its length.This damage likely occurred due to forceful manipulation of the cat6 during navigation of patient anatomy.The kinks throughout the proximal segment were likely incidental and may have occurred during packaging for return to penumbra.The guidewire mentioned in the complaint was not returned for evaluation.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure in the femoral artery using an indigo system aspiration catheter 6 (cat6).During the procedure, while attempting to retract a guidewire from a cat6, the cat6 grabbed the guidewire and it became stuck; therefore, the cat6 and guidewire were removed.The procedure was completed using infusion catheters.There was no report of an adverse effect to the patient.
|