An event regarding instability(varus/valgus laxity) and implant damage involving a custom insert was reported.The damage to the insert was confirmed through the images provided.Method & results: device evaluation and results: the reported device was not returned however photographs were provided for review.The photographs shows a recently explanted custom triathlon poly insert.The posterior of one of the two condyles shows delamination resulting in a crack of the polyethylene outer layer consistent with wear of the bearing surface most likely caused by posterior edge loading from the femoral component.On the lateral aspect of the bearing surface of one of the compartments, damage is observed, however, the cause of this damage cannot be determined.Damage consistent with explantation is also visible on the underside of the insert.Medical records received and evaluation: no medical records or x-rays were made available for evaluation.Device history review: device history review indicated the device was free from discrepancies when released.Complaint history review: not applicable as this is a custom device.Conclusions: the reported device was not returned however photographs were provided for review.The photographs shows a recently explanted custom triathlon poly insert.The posterior of one of the two condyles shows delamination resulting in a crack of the polyethylene outer layer consistent with wear of the bearing surface most likely caused by posterior edge loading from the femoral component.On the lateral aspect of the bearing surface of one of the compartments, damage is observed, however, the cause of this damage cannot be determined.Damage consistent with explantation is also visible on the underside of the insert.While the event could be confirmed through the explant images provided, the root cause could not be determined because the insufficient medical information was provided.Further information such as x-rays, medical records, return of device, patient history, histopathology report & follow-up notes are needed to investigate this event further.If additional information and/or device becomes available, this investigation will be reopened.
|