A customer reported on (b)(6) 2018, the stardust dnh str #278093, spring¿s handle suddenly broke while the surgeon was suturing the patient.Also during use suspect device, rb9344 yarargil micro dissector bayo, tip broke and fell in the patient¿s brain cavity.The surgeon retrieved the broken tip from the patient.It was noted there was a 15-minute delay of the surgery.The surgery was completed with other instruments and the patient was fine.This is one of two reports concerning the same patient and procedure using two integra devices.
|