BECTON, DICKINSON & CO., (BD) BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET W/ TUBING AND HOLDER; BLOOD COLLECTION TUBE
|
Back to Search Results |
|
Catalog Number 368652 |
Device Problems
Sticking (1597); Separation Failure (2547)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/05/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter phone#: (b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the tubing on a bd vacutainer® push button blood collection set with tubing and pre-attached holder was stuck together.When trying to pull it apart, holes were ripped in the tubing.The device was not used and there was no report of injury or medical interventions.
|
|
Manufacturer Narrative
|
Investigation summary: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.A review of the manufacturing records was completed for the incident lot and no issues were identified.Investigation conclusion: as no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.Root cause description: as there was no sample or photo available for evaluation, a root cause could not be determined.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
|