(b)(4).The customer, a biomedical engineer, evaluated the device and verified the reported issue.The customer confirmed that the hard paddles had a broken pin in the connector.The hard paddles were replaced which resolved the reported issue.The customer confirmed that the damaged hard paddles were disposed of.After observing proper device operation through functional and performance testing, the device was returned for use.The device was not returned to physio-control for evaluation.Device not evaluated by manufacturer.
|
The customer contacted physio-control to report that the charge button on their hard paddles was not functioning.The customer observed a broken pin on the connector of the hard paddles.As a result, defibrillation therapy would not be able to be delivered, if it were necessary.There was no patient use associated with the reported event.
|