Device evaluated by mfr: the returned product consisted of a watchman hemostasis valve in the unlocked position.Based on the potential root causes identified in the fmea, the seals, plunger, luer and port were microscopically and visually inspected.Inspection of the device presented no damage or irregularities.Since the devices used in the clinical event were not returned for analysis, a test mach1 guide catheter, a balloon catheter and wire were used for functional testing.The mach 1 was attached to the watchdog, then a balloon catheter with a wire was inserted into the watchdog and guide catheter.The end of the mach 1 with the balloon and wire was placed in a glass of water, then a syringe full of water was hooked up to the side port.The syringe was pulled and pushed and there were no air leaks when the distal seal is locked.The distal seal was unlocked and tested again and air does go into the syringe when the distal seal is unlocked.The investigation conclusion is user/use error as there was an act or omission of an act that resulted in a different medical device response than intended by the manufacturer or expected by the user.The watchdog directions for use includes the following instruction: if required to draw a blood sample attach a syringe to the manifold system or side port of the watchdog hemostasis valve and withdraw the blood.Before withdrawing blood, always close the distal seal by rotating the plunger in clockwise direction.(b)(4).
|