Model Number G8
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/16/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This report is being submitted due to a retrospective review conducted under capa-(b)(4).On (b)(4) 2017, fse arrived at the site to address the reported event.Fse replaced the tube for the line filter.The customer was subsequently able to run controls and verify results.No further issues were noted.No further action was required by field service.The most probable cause of the reported event was due to fault/ failure of the line filter tube.
 
Event Description
On (b)(6) 2017, the customer reported the peek fitting that attached to the top of the prefilter housing for their g8 analyzer had broken off.The customer stated that she was able to remove the broken fitting, but she was unable to locate a new peek compression fitting to replace the broken one.On (b)(4) 2017, a field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hba1c.There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba- koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7368244
MDR Text Key103693657
Report Number8031673-2018-04241
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/23/2018
Distributor Facility Aware Date06/16/2017
Device Age1 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/23/2018
Date Manufacturer Received06/16/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-