• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND VENA SEAL CLOSURE SYSTEM; AGENT, OCCLUDING, VASCULAR, PERMANENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VENA SEAL CLOSURE SYSTEM; AGENT, OCCLUDING, VASCULAR, PERMANENT Back to Search Results
Model Number SP-101
Device Problem Stretched (1601)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/26/2018
Event Type  malfunction  
Manufacturer Narrative
Two images were received: image 1 revealed a guidewire inserted into a blue dilator.The image also revealed the guidewire is also inserted into the catheter from the white transportation hoop.Image 2 revealed a damaged guidewire.The image shows tensile stretching of the guidewire resulting from tearing of the coating on guidewire.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the physician intended to use venaseal to treat the great saphenous vein as per ifu.It was reported the guidewire was advanced to the junction and the venaseal sheath and dilator was advanced over the wire to the junction.The wire was withdrawn from the sheath and it was noted that the wire was frayed.The wire was removed from the patient as a precautionary measure.The procedure was completed with another venaseal and no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENA SEAL CLOSURE SYSTEM
Type of Device
AGENT, OCCLUDING, VASCULAR, PERMANENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key7376755
MDR Text Key103660811
Report Number9612164-2018-00586
Device Sequence Number1
Product Code PJQ
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
P140018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2018
Device Model NumberSP-101
Device Lot Number44321
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/27/2018
Date Device Manufactured07/12/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-