• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, 10MM PEEK MONOPOLAR HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, 10MM PEEK MONOPOLAR HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250080618
Device Problem Device Packaging Compromised (2916)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/01/2018
Event Type  malfunction  
Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.(b)(4).
 
Event Description
It was reported that the insulation had been compromised.
 
Manufacturer Narrative
Gtin: (b)(4).The device manufacture date is not known.Alleged failure: cracked shaft insulation the failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be user misuse, excessive force, normal wear, or improper sterilization methods.The product was returned for investigation and the failure mode was confirmed and will be monitored for future reoccurrence.
 
Event Description
It was reported that the insulation had been compromised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 10MM PEEK MONOPOLAR HANDLE
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
ashley lower
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key7377708
MDR Text Key103796174
Report Number0002936485-2018-00280
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250080618
Device Lot Number1243097D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received03/01/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-