(b)(4).Device evaluated by mfr.: the complaint device was not received for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
Same case as mdr id#: 2134265-2018-02312.It was reported that the stent foreshortening occurred.The target lesion was located in the left iliac vein.A 18x60 mm wallstent® endoprosthesis stent was implanted for treatment.However, the stent was noted to be shortened and had to place additional 18x40 wallstent® endoprosthesis stent.The stent was successfully deployed; however, upon removal of the delivery system, resistance was encountered.The delivery system hung up on the sheath at the distal marker.The device was advanced further into the inferior vena cava and back through the sheath.During this time, the delivery system became stuck at the proximal marker.Eventually, the sheath and the delivery system was removed as a unit and the procedure was completed.No patient complications were reported and the patient was not negatively affected.
|