• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TULSA DENTAL PRODUCTS LLC X-SMART IQ MOTOR; CONTROLLER, FOOT, HANDPIECE AND CORD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TULSA DENTAL PRODUCTS LLC X-SMART IQ MOTOR; CONTROLLER, FOOT, HANDPIECE AND CORD Back to Search Results
Catalog Number XSIQEM1HP
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device was evaluated and found to be within specification.
 
Event Description
It was reported that an x-smart iq was set on recipicating mode but was moving in rotary motion.The device did not properly auto-reverse and a file separated at the apex of a patient's tooth.The doctor left the broken portion of the file in the root canal and incorporated it into the filling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-SMART IQ MOTOR
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
TULSA DENTAL PRODUCTS LLC
608 rolling hills drive
johnson city TN 37604
Manufacturer (Section G)
ADVANCED TECHNOLOGY RESEARCH (A.T.R.) S.R.L.
via san donato 1
pistoia, 51100
IT   51100
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key7386090
MDR Text Key104036159
Report Number2320721-2018-00014
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 03/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberXSIQEM1HP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/01/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-