• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE MAHWAH CARDIOSAVE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE MAHWAH CARDIOSAVE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Invalid Sensing (2293)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/12/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed per company standard operating procedure since the device manufacture date is greater than one year from the event date.A getinge service territory manager (stm) evaluated the iabp and was unable to duplicate the reported issue.The stm tested the fiber optic module with calibrated chamber which passed to specifications.The iabp passed all functional and safety tests per factory specifications and was returned to customer and cleared for clinical use.
 
Event Description
The customer reported that while the cardiosave intra-aortic balloon pump (iabp) was used on a patient, the iabp had a fiber optic - amplitude of arterial wave that was clipped, so the customer tried all options and then swapped pumps.No patient harm or serious injury was reported, and there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7394924
MDR Text Key104423801
Report Number2249723-2018-00545
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 04/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-