• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problems Charred (1086); Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/26/2018
Event Type  malfunction  
Event Description
It was reported that the patient¿s wife called to report that the transmitter would not power up.The prongs were disconnected and black charring was noted on the left contact; the green contact was damaged.The transmitter was replaced.
 
Manufacturer Narrative
The field complaint of the green strips having charring and the transmitter would not power up was verified.After removing the prongs, burn marks were observed and upon opening the ac power adapter, burnt components were observed on the main circuit board.A burnt mark on the inner casing of the ac power adapter was also observed.Upon opening the transmitter; burnt modem circuit components were observed on the main pcb.It was concluded that the root cause of no power issue was the burnt damage that the ac adapter¿s main pcb and the transmitter main pcb sustained through the ac electrical power outlet.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7395565
MDR Text Key104284340
Report Number2017865-2018-05059
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Patient Family Member or Friend
Type of Report Initial,Followup
Report Date 04/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Device Catalogue NumberEX1150
Other Device ID Number05414734504799
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received04/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/17/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-