On (b)(6) 2018, a 19mm sjm regent heart valve was implanted.The leaflet function was tested following implant using the silicon tester.Due to the position of aorta, the valve was not rotated.The aorta was closed and an echo was performed showing an abnormal result.Before sternal rewire, the aorta was reopened and one leaflet was noted to be dislodged.The leaflet was fractured and broken in two pieces, one located by the suturing cuff and the largest piece in the lv.Both pieces were removed from the patient.A 21mm medtronic valve (model unknown) was implanted.The patient is reported to be stable.
|
The reported event of a fractured leaflet was confirmed.The majority of the leaflet was returned in two pieces, with the left ear remaining missing.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.There was no evidence of material defect in the carbon coating that may have caused or contributed to the dislodged leaflet or orifice damage.However, the damage was consistent with some external force applied to the valve which overstressed the carbon material.The cause of the reported event remains unknown.
|