(b)(4).Batch # p94c67, p9472k, p9477w.Device analysis: only what appears to be a half of a ligamax5 clip was returned for analysis.Upon visual inspection of the clip, it was found to be broken in two parts and only one was returned.As the device was not returned for analysis, no further testing could be performed.As per the instructions for use "a thorough understanding of the principles and techniques involved in laser, electrosurgical, and ultrasonic procedures is essential to avoid shock and burn hazards to both patient and medical personnel and damage to the device or other medical instruments." the batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|