During preparation for a thrombectomy procedure, the hospital staff found that the penumbra system 5max reperfusion catheter (5maxc) was broken approximately 40cm from the proximal hub upon removal from the packaging.The damage to the 5maxc was found prior to use and therefore, it was not used in the procedure.The procedure was completed using a new 5maxc.
|
Results: the 5maxc was fractured approximately 82.0 cm from the hub.Conclusions: evaluation of the returned 5maxc revealed a fractured device.If the 5maxc is forcefully removed from the packaging hoop at extreme angles or otherwise mishandled during preparation, damage such as a kink may occur.If a kink is further manipulated or mishandled, the kink may become fractured.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|