The customer observed multiple falsely elevated total bilirubin results while using the clinical chemistry total bilirubin reagents.The following data was provided.Sid (b)(6) initial 1.636 mg/dl, repeat 0.509 mg/dl, sid (b)(6) initial 1.807 mg/dl, repeat 0.717 mg/dl, sid (b)(6) initial 1.646 mg/dl, repeat 0.493 mg/dl, sid (b)(6) initial 1.795 mg/dl, repeat 0.674 mg/dl, sid (b)(6) initial 1.582 mg/dl, repeat 0.443 mg/dl, sid (b)(6) initial 1.581 mg/dl, repeat 0.431 mg/dl, sid (b)(6) initial 2.088 mg/dl, repeat 1.045 mg/dl, sid (b)(6) initial 1.711 mg/dl, repeat 0.578 mg/dl, sid (b)(6) initial 1.654 mg/dl, repeat 0.512 mg/dl, sid (b)(6) initial 1.778 mg/dl, repeat 0.66 mg/dl, sid (b)(6) initial 1.835 mg/dl, repeat 0.781 mg/dl, sid (b)(6) initial 1.648 mg/dl, repeat 0.528 mg/dl, sid (b)(6) initial 1.623 mg/dl, repeat 0.453 mg/dl, sid (b)(6) initial 1.605 mg/dl, repeat 0.472 mg/dl, sid (b)(6) initial 1.474 mg/dl, repeat 0.34 mg/dl, sid (b)(6) initial 1.699 mg/dl, repeat 0.592 mg/dl, sid (b)(6) initial 1.548 mg/dl, repeat 0.427 mg/dl, sid (b)(6) initial 1.734 mg/dl, repeat 0.608 mg/dl, sid (b)(6) initial 1.425 mg/dl, repeat 0.278 mg/dl, sid (b)(6) initial 1.756 mg/dl, repeat 0.625 mg/dl, sid (b)(6) initial 1.648 mg/dl, repeat 0.505 mg/dl, sid (b)(6) initial 1.373 mg/dl, repeat 0.203 mg/dl, sid (b)(6) initial 1.412 mg/dl, repeat 0.262 mg/dl, sid (b)(6) initial 1.335 mg/dl, repeat 0.193 mg/dl, sid (b)(6) initial 1.915 mg/dl, repeat 0.828 mg/dl, sid (b)(6) initial 1.383 mg/dl, repeat 0.219 mg/dl, sid (b)(6) initial 1.909 mg/dl, repeat 0.802 mg/dl, sid (b)(6) initial 2.137 mg/dl, repeat 1.087 mg/dl, sid (b)(6) initial 1.603 mg/dl, repeat 0.469 mg/dl, sid (b)(6) initial 1.695 mg/dl, repeat 0.564 mg/dl, sid (b)(6) initial 1.508 mg/dl, repeat 0.337 mg/dl, sid (b)(6) initial 1.896 mg/dl, repeat 0.804 mg/dl, sid (b)(6) initial 1.638 mg/dl, repeat 0.514 mg/dl, sid (b)(6) initial 1.494 mg/dl, repeat 0.321 mg/dl, sid (b)(6) initial 2.027 mg/dl, repeat 0.972 mg/dl, sid (b)(6) initial 1.705 mg/dl, repeat 0.594 mg/dl, sid (b)(6) initial 1.658 mg/dl, repeat 0.53 mg/dl, sid (b)(6) initial 1.674 mg/dl, repeat 0.564 mg/dl, sid (b)(6) initial 1.738 mg/dl, repeat 0.608 mg/dl, sid (b)(6) initial 1.615 mg/dl, repeat 0.481 mg/dl, sid (b)(6) initial 1.695 mg/dl, repeat 0.557 mg/dl, sid (b)(6) initial 1.668 mg/dl, repeat 0.54 mg/dl, sid (b)(6) initial 1.689 mg/dl, repeat 0.568 mg/dl, sid (b)(6) initial 1.394 mg/dl, repeat 0.233 mg/dl, sid (b)(6) initial 2.092 mg/dl, repeat 1 mg/dl, sid (b)(6) initial 1.953 mg/dl, repeat 0.87 mg/dl, sid (b)(6) initial 1.471 mg/dl, repeat 0.288 mg/dl, sid (b)(6) initial 1.778 mg/dl, repeat 0.679 mg/dl, sid (b)(6) initial 1.742 mg/dl, repeat 0.627 mg/dl, sid (b)(6) initial 1.782 mg/dl, repeat 0.658 mg/dl, sid (b)(6) initial 2.909 mg/dl, repeat 1.916 mg/dl, sid (b)(6) initial 1.44 mg/dl, repeat 0.288 mg/dl, sid (b)(6) initial 1.752 mg/dl, repeat 0.623 mg/dl, sid (b)(6) initial 1.709 mg/dl, repeat 0.571 mg/dl, sid (b)(6) initial 1.957 mg/dl, repeat 0.867 mg/dl, sid (b)(6) initial 3.045 mg/dl, repeat 2.082 mg/dl, sid (b)(6) initial 1.959 mg/dl, repeat 0.854 mg/dl, sid (b)(6) initial 1.636 mg/dl, repeat 0.493 mg/dl, sid (b)(6) initial 1.825 mg/dl, repeat 0.733 mg/dl, sid (b)(6) initial 2.491 mg/dl, repeat 1.467 mg/dl, sid (b)(6) initial 1.45 mg/dl, repeat 0.287 mg/dl, sid (b)(6) initial 2.163 mg/dl, repeat 1.122 mg/dl no impact to patient management was reported.
|