• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC SELENIA MAMMOGRAPHY SYSTEM; FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC SELENIA MAMMOGRAPHY SYSTEM; FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Back to Search Results
Model Number SEL-00030
Device Problem Unintended Movement (3026)
Patient Problem Not Applicable (3189)
Event Date 02/12/2018
Event Type  malfunction  
Event Description
It was reported that during a patient exam the paddle came down on a patient uncommanded and the paddle was not able to be moved back upward.There was no patient injury reported, which was confirmed with the site.It was determined that the angulation drag brake needed to be replaced and once this was completed the system was working as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA MAMMOGRAPHY SYSTEM
Type of Device
FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
36 & 37 apple ridge road
danbury CT 06810
Manufacturer Contact
kristin fornieri
36 & 37 apple ridge road
danbury, CT 06810
2037318491
MDR Report Key7403616
MDR Text Key104808940
Report Number1220984-2018-00095
Device Sequence Number1
Product Code MUE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Radiologic Technologist
Type of Report Initial
Report Date 03/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model NumberSEL-00030
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-