(b)(4).Batch # p93n1v.Device analysis: the analysis results found that the er320 device was returned with no damage in the external components with a clip in the jaws; the clip was removed in order to inspect the jaws and they were found with no damage.In addition, the tyvek was returned along with the instrument.In an attempt to replicate the reported incident, the device was tested for functionality.During the analysis, the device was cycled and it fed, retained and formed the remaining 19 clips as intended.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, no anomalies were found.As the device was found to be fully functional, it could not be determined what may have caused the reported incident.No conclusion could be reached as to what may have caused the reported incident.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|