MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*MCA SMALL APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*MCA SMALL APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number MCS20
Device Problem Component Falling (1105)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/22/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
 
Event Description
There were multiple issues during a thyroidextomy: 2 at the same time, then no clip that released upon firing, clips that fell out, clips that didn't hold on well.No patient consequence.
 
Manufacturer Narrative
(b)(4).Device analysis: the analysis results found that the mcs20 device was returned with no damage in the external components.In addition, the tyvek was returned along with the instrument.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled and it fed and formed the remaining 5 clips as intended.However, the lock out mechanism was noted to be non-functional.In order to evaluate the condition of the internal components, the device was disassembled and the lock out spring was noted out of position.No conclusion could be reached as to what may have caused the lockout spring to be out of position.The lot history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the packaging process.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7406782
MDR Text Key105258035
Report Number3005075853-2018-09013
Device Sequence Number1
Product Code GDO
UDI-Device Identifier20705036002496
UDI-Public20705036002496
Combination Product (y/n)N
PMA/PMN Number
K820837
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2022
Device Catalogue NumberMCS20
Device Lot NumberP4T72Z
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received04/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-