Model Number TJF-160VR
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Information (3190)
Event Date 03/07/2018
Event Type  malfunction  
Manufacturer Narrative
The subject device has not been returned to omsc.But was returned to olympus (b)(4).The subject device was sent to a third party laboratory for additional microbiological testing.In the additional test, staphylococcus coagulase negative (1 cfu/100ml) were detected from the subject device.The testing result cleared french guideline.Omsc reviewed the manufacturing history of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that as a result of routine microbiological testing by the user facility, the subject device tested positive for aeromonas hydrophila/caviae (>100cfu/scope).The subject device had been manually reprocessed with peracetic acid.There was no report of infection associated with this report.
 
Manufacturer Narrative
This supplemental report is submitting to correct "device product code".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
MDR Report Key7410580
MDR Text Key105393639
Report Number8010047-2018-00597
Device Sequence Number1
Product Code FDT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-160VR
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/28/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-